DESCRIPTION:Speakers: Yongbo Huang (Guangxi University)\nA 20 kton liquid 
 scintillator (LS) detector was designed in the Jiangmen Underground Neutri
 no Observatory (JUNO) for multiple physics purposes. In order to determine
  the neutrino mass ordering\, JUNO needs an excellent energy resolution of
  3% /√E(MeV) and an accuracy of the energy scale at 1% level or better. 
 On one hand\, a comprehensive calibration system is designed\, deploying m
 ultiple radioactive sources in various locations inside/outside of the cen
 tral detector (CD)\, including Auto Calibration Unit (ACU)\, Cable Loop Sy
 stem (CLS)\, Guide Tube Calibration System (GTCS)\, and Remotely Operated 
 Vehicle (ROV). On the other hand\, the strategy of the JUNO calibration sy
 stem has been developed and optimized based on Monte Carlo simulation resu
 lts. Following this calibration strategy\, we expect to achieve an accurat
 e and comprehensive understanding of the energy resolution\, energy non-li
 nearity and non-uniformity responses of the JUNO detector. In this talk\, 
 the details of the JUNO calibration system design and calibration strategy
  will be presented.\n\nhttps://indico.tlabs.ac.za/event/112/contributions/